  Summary of M05  (to CPPM) 
  ============== 

  cycle  absorbers    absorbers     Comments
         delivered      good       
                    (Preliminar)  

  1       10G+4P       9G+4P       b0595_03 shows raugh area 60x40x0.1-0.2 
                                   in approx. middle of absorber ... 
                                   this absorber has to be checked carefuly 
                                   before using it. I declare it not OK for 
                                   use (for the time being).  
  
  2       10G+4P      10G+4P       b0597_02,b0606_06,b0697_09: important hits 
                                   with deformation approx. 15x5 mm2
                                   at the most obtuse tip of those absorbers 
                                   are reported at QC sheets.
 
                                   s181_04: important hit at obtuse edge 
                                   close to wave 8 (15x40). Lead-SS unglued  
                                   at both sides. 
 
                                   Has Fiber people tried to repair those 
                                   absorbers ? Can those absorbers be used ? 
                                   Let us assume yes, but check them 
                                   carefully. 
                   
  3       10G+4P      10G+4P       b0345_06_m05c03: an important hit with 
                                   deformation 15x5 mm2 at the acute edge 
                                   is reported at QC sheet. Has Fiber people
                                   tried to repair it ? Can be that absorber 
                                   used ? Let us assume yes, but check it 
                                   carefully

  4       10G+4P      10G+4P       s207_01: at acute edge half-wave 8 present
                                   an important hit of 60x2 mm2. Let us assume 
                                   that the absorbers can be used, but check it
                                   carefully.  

                                   b0616_04: thicknesses at obtuse edge, first 
                                   8 waves are significantly higher than 
                                   nominal. In 3D measurements all normal 
                                   except, maybe a couple of angles at the 
                                   obtuse side of the absorber.
                                   One should check it carefully
 
  5        10G+4P      10G+4P      s211_03: shows larger thicknesses at 
                                   obtuse/back than normal (from basic QC). 
                                   Check the absorber carefully.
 
  6        10G+4P      10G+4P
  7        10G+4P      10G+4P
  8        10G+4P      10G+4P
  9        10G+4P      10G+4P
 10        10G         10G



       Total absorbers delivered: 100G+36P  
       Total abs. good (prelim):   99G+36P